AC/DC withstand voltage insulation impedance tester
【Product Name】AC/DC withstand voltage insulation impedance tester
【Product Category】线束综合测试设备
【Popular Products】577
【Updated】2023/3/21 9:53:23
Details

AC/DC withstand voltage insulation impedance tester

brief introduction

The AC and DC withstand voltage/insulation testers use linear amplification technology to generate low distortion sine wave output, making the testing more accurate. The use of a 4.3 inch color LCD display screen and a user-friendly interactive operation mode make it clearer and easier for users to understand. The test results and test conditions are displayed simultaneously, making users feel at ease with the test conclusion. This instrument can be widely used in fields that require safety testing.

Performance characteristics

4.3 inch color liquid display
Linear, low distortion sine wave output
5 sets of program settings and memory functions
Voltage ramp up setting function
Unique arc detection function for more accurate defect identification
Equipped with AC/DC withstand voltage/insulation resistance testing function
Equipped with AC/DC withstand voltage testing function

Model 7122D
AC withstand voltage test output voltage range:0-5000V,0-12mA,resolution ratio:10V
accuracy:±(2% set value+5V)
output frequency 50Hz/60Hz
Output waveform Sine wave chord, distortion<2%
Upper limit setting range:0.10-12.00mA,resolution ratio:0.01mA
accuracy:±(2% set value+5V)
Lower limit setting range:0.00-12.00mA,resolution ratio 0.01mA
accuracy:±(2% set value+5V)
ARC Detector 1-9 Level
DC withstand voltage test Output voltage range:0-6000V,
resolution ratio 10V,accuracy:±(2% set value+5V)
OUTPUT RIPPLE <5%(6KV/5mA)
Testing under linear load
Upper limit setting range:0.02-5.00mA,resolution ratio  0.01mA  accuracy:±(2% set value+5V))
Lower limit setting range:0.00-5.00mA,resolution ratio  0.01mA  accuracy:±(2% set value+5V)
Automatic discharge 200mS
control interface Input: Test, RESET
Output: Test Passed, Test Failed, Test In Progress
Ramp up setting 0.1-999.9 seconds
Test time Range: 0.5-999.9 seconds (0=continuous testing)
Output method of test results Buzzer, LCD display reading, control interface status output
Memory group 5 sets of test condition memory, each with 3 test steps to choose from
display 4.3 inch color LCD display
Insulation resistance test output voltage Range: DC100-1000V, resolution 10V,
Accuracy: ± (2% set value+5V)
Resistance range Range: 1-1000M Ω, (4-digit display, automatic shifting) Accuracy: ± (3% reading value+2 words) Voltage ≥ 500V ± (7% reading value+2 words) Voltage<500V
Upper limit setting Same resistance range, spacing of 1M Ω
Lower limit setting Same resistance range, spacing of 1M Ω